Skip to main content

Table 2 Optimized sleep parameters determined by the FS-760 actigraph

From: Validity of an algorithm for determining sleep/wake states using FS-760 in school-aged children

 

PSG

FS-760

 

Before optimization

After optimization

Mean±SE

Mean±SE

t

p

icc

p

Mean±SE

t

p

icc

p

Sleep latency (min)

10.00 ± 9.83

7.85 ± 9.16

-1.292

0.199

0.649

<0.001

9.38 ± 11.61

-0.326

0.745

0.654

<0.001

Wake after sleep onset (min)

26.52 ± 29.69

44.37 ± 34.27

3.173

0.002

0.491

0.001

24.65 ± 27.92

0.960

0.339

0.613

<0.001

Total sleep taime (min)

380.31 ± 32.24

364.62 ± 37.27

-2.567

0.011

0.524

<0.001

385.97 ± 34.96

-0.371

0.711

0.625

<0.001

Sleep efficiency (%)

90.55 ± 7.68

86.81 ± 8.87

-2.567

0.011

0.524

<0.001

91.9 ± 8.32

0.960

0.339

0.613

<0.001